Is there a need for individualized adjustment of electrode belt position during EIT-guided titration of positive end-expiratory pressure?
在eit引导PEEP滴定时,是否需要个体化调整电极带的位置?


杂志名称:Physiological measurement

IF(2022-2024年):3.2

发表时间:2022 Jun.


Abstract


Objective. The aim of the present study was to evaluate the variation of tidal volume-to-impedance ratio (VT/ZT) during positive end-expiratory pressure (PEEP) titration with electrical impedance tomography (EIT) measurement.


Approach. Forty-two patients with acute respiratory distress syndrome were retrospectively analyzed. An incremental and subsequently a decremental PEEP trial were performed with steps of 2 cm H2O and duration of 2 min per step during volume-controlled ventilation with decelerating flow. EIT measurement was conducted in the 5th intercostal space andVTwas recorded simultaneously. The variation ofVT/ZT(RatioV) was defined as the changes in percentage to average ratio per cm H2O PEEP change. A z-score > 1 was considered as a significant variation and an implication that the measurement plane was inadequate.


Main results. TheRatioVof 42 patients was 1.29 ± 0.80%·cm H2O-1. A z-score of 1 corresponded to the variation of 2.09%·cm H2O-1. Seven patients (16.7%) had a z-score > 1 and showed either positive or negative correlation between the volume-to-impedance ratio and PEEP.


Significance. Electrode placement at 5th intercostal space might not be ideal for every individual during EIT measurement. Evaluation of volume-to-impedance ratio variation is necessary for patients undergoing maneuvers with wide alteration in absolute lung volume.


https://pubmed.ncbi.nlm.nih.gov/35617942/